CONDUCTIVITY
Range
Resolution
Manual
Automati c
Automatic (mS/cm)
Accuracy
Calibration
RESISTIVITY
Range
(depending on measurement setup)
Resolution
Calibration
TDS (Total Dissolved Solids)
Range
Resolution
Manual
Automati c
Automatic ppt (g/L)
Accuracy
Calibration
1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm
0.01 mS/cm from 10.00 to 99.99 mS/cm
0.1 mS/cm from 100.0 to 400.0 mS/cm
0.001 mS/cm from 0.000 to 9.999 mS/cm
0.01 mS/cm from 10.00 to 99.99 mS/cm
0.1 mS/cm from 100.0 to 400.0 mS/cm
±1 % of reading or ±1 µS/cm whichever is greater
Automatic single point, with 6 standard solutions
(84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm,
80.0 mS/cm, 111.8 mS/cm) or custom point
Based on conductivity or salinity calibration
(the maximum value depends on the TDS factor)
0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L)
1 ppm (mg/L) from 0 to 9999 ppm (mg/L)
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L)
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L)
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L)
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
±1 % of reading or ±1 ppm (mg/L) whichever is greater
Based on conductivity or salinity calibration
0 to 200 mS/cm
(absolute EC up to 400 mS/cm)
1 µS/cm from 0 to 9999 µS/cm
0 to 999999 Ω·cm;
0 to 1000.0 kΩ·cm;
0 to 1.0000 MΩ·cm
Depending on resistivity reading
0 to 400000 ppm (mg/L);
1 ppm (mg/L); 0.001 ppt (g/L);