Running A Control Solution Test - EPS Bio Technology EasyMax NG User Manual

Next generation (ng) self-monitoring blood glucose system
Table of Contents

Advertisement

Running a Control Solution Test

You need the meter, a test strip, and control solution.
1. Put a test strip into
the meter in the
direction of the
arrow.
Note:
Setting the meter in control solution mode will prevent the control solution reading from being
added to your memory and averages.
2. The meter turns on automatically and the strip
icon appears.
Press
(Left button) to change the testing mode to
Control Solution and the
icon will flash.
3.
Place the meter on a flat
surface, like a table.
4.
Remove the control solution
bottle cap and wipe the tip of
the bottle with a tissue.
5.
Squeeze the bottle until a
tiny drop forms at the tip of
the bottle and place the
solution on top of bottle cap.
19

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EasyMax NG and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents