Cleaning And Disinfection Frequency
At least 9 cycles of cleaning per day and 1 cycle of disinfection per week will allow the user to use the device for
4 years. The meter can sustain 20,000 cycles of cleaning and disinfection and the lancing device can sustain
14,600 cycles of cleaning and disinfection.
Cleaning Frequency: 9 cycles/day X 365 days/year X 4 year product life = 13,140 cycles of cleaning
Disinfection Frequency: 1 cycle/week X 52 weeks/year X 4 year product life = 208 cycles of disinfections
< 20,000 validated cycles of cleaning and disinfection for the meter
< 14,600 validated cycles of cleaning and disinfection for the lancing device
4 year product life is for properly cleaning and disinfection. After 4 years, the meter must be replaced
with a new meter.
37
Need help?
Do you have a question about the EasyMax NG and is the answer not in the manual?