Step 1: Cleaning Instruction:
and bodily fluids must be thoroughly cleaned from surfaces and objects before disinfection with a germicidal
wipe. Open, unfold and use the first germicidal wipe to remove heavy soil.
Step 2: Disinfection Instruction:
thoroughly wet all meter surfaces, including the strip port and the connection port. Treated area must remain
visibly wet for a full 2 minutes. Use additional wipe(s) if needed to ensure a continuous 2 minute wet contact
time. Let the device air dry for 30 seconds.
Do:
Make sure the meter is turned off during cleaning and
disinfection.
Keep test strip vial(s) tightly closed when performing
cleaning and disinfection procedures; the fumes from
the disinfectant may affect the performance of the
strips.
After cleaning and disinfection, please perform a
physical appearance and performance check of
devices.
Always wear gloves when cleaning and testing.
Note:
After disinfection, users' gloves should be removed and hands should be thoroughly washed with
soap and water before proceeding to the next patient.
Clean first to remove any heavy soil left on the surface of the meter. All blood
After cleaning, and before using the meter again, unfold a new wipe and
Do Not:
Get any moisture into the test strip slot.
Spray any cleaning solution directly onto
the meter.
Put the meter under water or liquid.
Pour liquid onto the meter.
34