Calibrating The Scanner; Replacing The Separation Pad - Mustek iDocScan D20 User Manual

High performance document scanner
Hide thumbs Also See for iDocScan D20:
Table of Contents

Advertisement

MAINTENANCE

Calibrating the Scanner

When scanned image's colors appear not to match the colors of the
original, or colors appear in the white areas of the images, calibration
helps to correct these problems by compensating variations in the
scanner components that happens over time.
1. Turn on the power of the scanner.
2. Press Start > All Programs > iDocScan
D20 > Calibration Utility to open the
calibration screen.
3. Remove the plastic foil on both
sides of the Calibration Sheet.
4. Insert the Calibration Sheet into the
ADF tray with the arrow mark facing
up and forward.
5. Press "Calibrate" to start calibrating.
The Calibration Sheet will be
gradually fed through the ADF.
 A progress bar will appear
indicating the calibration process.
 Mechanical sounds produced during
the calibrating process are normal.
6. Press "Close" after "the calibration is
done" window appears.
20

Replacing the Separation Pad

The Separation Pad in the ADF part is a consumable item. After approx.
30000 times of scanning, it may become worn. Replace the Separation
Pad when it is worn, damage or misfeeding continues to occur after
cleaning. A spare Separation Pad is included with the scanner, please
replace it as following steps.
1. Turn off the power, and then lift open the ADF Cover.
2. Press both springs on each side of the Separation Pad, pull the pad
backwards and lift it off the ADF Cover.