(5) Start water cooling and water cooping in operation
・ Confirm the water is flowing before starting the camera cooling and that the camera
does cool.
・ Keep 1.0 L/min flow rate or 0.25 MPa water pressure for water circulation.
・ Do not stop coolant while the camera is working.
(6) Cooling water hose
The hose has a blue hose (Internal diameter: 4 mm / External diameter: 6 mm) and a
gray hose (Internal diameter: 8 mm / External diameter: 13.5 mm). (Figure 8-1)
If the hose size on circulating water cooler is same as blue hose, remove gray hose
from the joint part. The gray hose can be removed when blue hose is pulled with
pushing the button of the joint on gray hose. (Figure 8-2)
Gray hose
Figure 8-1
(7) Connection of Cooling water hose
CAUT ION
・ Stop water circulation when connecting / disconnecting the hose, and turn off the
power of the camera and the circulating water cooler.
・ Confirm that cooling water stops.
・ Prepare water absorption sheet (such as Waste, Towel or so) and catch pan in order
to avoid water drop or water splash.
(8) Deterioration of Cooling water hose
Replace the water hose with a new one whenever it cannot keep the Keep 1.0 L/min
flow rate or 0.25 MPa water pressure for water circulation due to the hose deterioration.
Blue hose
Follow the instruction in Section 8-2 [CONNECTION OF WATER
COOLING
HOSES]
WATER-COOLING HOSES] to connect / disconnect the hose.
C11440-22C Instruction manual_Ver.1.4
Joint
Figure 8-2
and
Section
8-3
[DISCONNECTION
Button
OF
17