19.0 hM3000 settings dialogue box

The procedure for configuring a HM3000 is structured in a logical sequence
so that the HM 3000 is configured in the correct order. For example, in the
"HM3000 Settings" dialogue box (shown below) the user must "Configure
modules" prior to being able to "Configure channels" and "Configure
outputs" (these buttons are greyed out so they cannot be accessed until you
have configured a module). FloCom
the entire configuration process to minimise user error.
The essential steps to configure your HydroMace 3000 are listed below.

General settings

Sensor power
Configure modules (Chapter 20.0)
Configure channels (Chapter 21.0)
Configure outputs (Chapter 22.0)
Configure FloSI (Chapter 23.0)
View configuration
essential Steps to configure your hydroMace 3000
1. Edit General Settings
2. Add Modules
3. Add Channels
4. Edit Channels
5. Add Outputs (where required)
6. Edit Outputs (where required)
7. Configure FloSI (where required)
8. Apply Settings
WarNINg: Once Steps 1 - 7 have been completed, apply all
settings by clicking "apply" from the "hM3000 settings"
dialogue box as shown above.
32
SeCtion b SoFtware Manual
uses this logical sequence throughout
+
Reset configuration
Chapter 19.1
Chapter 20.1
Chapter 21.1
Chapter 21.3
Chapter 22.1
Chapter 22.2
Chapter 23.0
RetuRn to toC
19.1 general settings
1. From the main menu click "Settings" and the "HM3000 Settings" dialogue
box will appear as shown above.
2. Click "Change" and the "General Settings" dialogue box will appear as
shown below.
Site Id
The "Site Id" is a text field used
to identify the location of the
device. When data is retrieved
from the device, it will be
identified by this name.
For example, the "Site Id" might
be the name of the treatment
plant or license number.
Measurement and
Logging Interval
The measurement interval is the
time at which the device 'wakes up'
to measure/calculate the values on
each of the configured channels.
Data is logged in the data file ONLY
at the logging interval.
The measurement interval will always be shorter than or equal to the
logging interval.
In the case of the measurement interval being shorter than the logging
interval, then the data logged will be the average of those values
calculated per measurement interval.
3. Click "Apply" and the "HM3000 Settings" dialogue box will re-appear.
NOte: the logging interval can only be a multiple of the
measuring interval.
RetuRn to toC
NOte: the instrument's
"Site Id" is required before
the hM 3000 configuration
can be applied successfully
MaCe HydRoMaCe 3000 Product Manual
33