Configuring Flosi Card - Mace HydroMace 3000 Product Manual

Weatherproof monitoring and telemetry platform for remote locations mace hydromace 3000
Table of Contents

Advertisement

22.2 editing outputs
Continued
-
22.2.1 Editing a 4-20mA output
Value at 4mA
Value at 20mA
Value at 4ma
The value of the "Source" channel at which the "4-20mA output" will be set to
4mA.
Value at 20ma
The value of the "Source" channel at which the "4-20mA output" will be set to
20mA.
averaging
The user can elect to output an averaged value on the "4-20mA output" by
checking the box. When checked, the "4-20mA output" will only be updated
at the HM 3000 logging interval otherwise it is updated at each measurement
interval. In the case of the measurement interval being shorter than the
logging interval, then the "4-20mA output" will be the average of those values
calculated per measurement interval.
Click "apply" to save the settings then click "exit" to return to the
"Configure outputs" dialogue box.
22.2.2 Editing a pulse output
Volume increment
Volume increment
The incremental value at which a pulse will be sent. In the example above, a
pulse will be sent whenever an increment of 10 Megalitres is totalised.
NOte: a maximum of 10 pulses per second may be sent per measurement
interval. For example, during a measurement interval of three minutes
(180 seconds) a maximum of 1,800 pulses can be sent. If the hM 3000 has
not sent all pulses by the time of the next measurement, pulses WILL
be lost.
Click "apply" to save the settings then click "exit" to return to the
"Configure outputs" dialogue box.
74
SeCtion b SoFtware Manual

23.0 Configuring FloSI Card

The Flo Series Serial Interface (FloSI) is an optional upgrade module for
HydroMace 3000 devices. The FloSI provides unconditional polling access to
Averaging
the most recent user configured measurement results of the HydroMace 3000.
NOte: Only one FloSI per hydroMace 3000 device is supported
RS-232, RS-485 and SDI-12, interfaces are available from the
HydroMace 3000 with the use of a FloSI. The FloSI enables users to request the
last measurement from HM 3000 without the complexity of communicating
via the host communications port.
The FloSI cannot be used to download or configure a HM 3000, and only one
serial protocol may be used at one time.
The FloSI requires no routine maintenance and has no user serviceable
components.
Configuration of the FloSI requires MACE Flocom+ (Version 1.1.0.12 or later)
software and successful PC communications with a MACE HydroMace 3000
device (Firmware Version 1.00.4 or later). The FloSI configuration settings are
available from the "HM3000 settings" dialogue box.
23.1 Configuring a FloSI
1. Establish communications with the device as per Chapter 17 - Connecting
to the device.
2. Ensure at least one module and one channel is configured as per
Chapters 20 - 21. Check the FloSI enabled check box.
3. Click "Configure FloSI" and the
"FloSI settings" dialogue box will
appear.
FloSI enabled check box
RetuRn to toC
RetuRn to toC
MaCe HydRoMaCe 3000 Product Manual
75