SECTION 19—HOUR METER
SECTION 19—HOUR METER

Replacing Hour Meter

DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
NOTE: For this procedure, refer to FIGURE 19.1.
1. Unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 22.
3. Remove control panel. Refer to Removing Control Panel on page 52.
4. Remove and label the two hour meter connectors and wires connected to back of hour
meter.
5. Spread retaining clips on hour meter housing that secure hour meter to control panel.
6. Remove hour meter by pushing meter through front of control panel.
7. Install new hour meter reversing STEPS 4-6.
8. Reinstall control panel. Refer to Removing Control Panel on page 52.
9. Reinstall cabinet. Refer to Removing Cabinet on page 22.
Hour Meter
Hour Meter
Housing
Hour Meter
Connectors
Wires
Hour Meter
Back of
Control Panel
Retaining Clips (Hour
Meter Housing)
FIGURE 19.1 Replacing Hour Meter
63
Part No 1154245
Perfecto
2