Troubleshooting flowcharts

Migration

Patient had MRI
Yes
Re-counsel
Remove & reinsert
OR
Reinsert device
Remove to
if possible
insert deeper
Migration
causing compromised sound quality,
comfort, visibility or prevents extended
wear use of Lyric2
Query Patient
(did he/she manipulate?)
May need to insert
new device
NO
Re-counsel
Jaw movement moving
Re-evaluate insertion
device?
depth, resize, check
(check under microscope
insertion depth
inspection)
Reinsert
(eventually with different size or insertion depth)
If still migrates
Contraindicate