20

SoundLync instruction

The SoundLync serves as a remote control for Lyric2
and enables the patient to remove the device if
necessary. This instruction will help you to counsel your
patients regarding use of the SoundLync. The illustration
on page 6 shows the different components of the
SoundLync.
Make sure your patient practices and understands all of
the SoundLync functions and is able to remove the
device if necessary.
The SoundLync enables the patient to switch the device
into three different modes (ON, SLEEP, OFF) and to
change the volume.
Before using the SoundLync, the keychain housing needs
to be unscrewed to expose the black magnet. To change
the device settings, the magnet should be close to the
microphone protection to trigger the change. Every
change of the device is confirmed by beeps or beep
sequences.
Toggle between ON and SLEEP-Mode
If Lyric2 is ON, briefly holding the magnet close to the
device then removing the magnet switches it to SLEEP-
mode (2 beeps). Briefly holding the magnet close to the
device again and then removing it switches Lyric2 back
ON (1 beep).
Change Volume
To change volume, first place Lyric2 in SLEEP- mode (2
beeps). Switch device back ON (1 beep) and continue
holding the magnet close to the device. After a pause of
2-3 seconds, the device begins to cycle through the
volume steps (soft to loud) starting at level one until the
MaxVol-Level is reached. Each level is indicated by a
short beep that increases in intensity as the volume
levels increase. Lyric2 will continue to cycle through the
volume levels until the magnet is removed at the desired
level. The MaxVol-level is set in the fitting software.
Remember
Volume-cycle always starts at level 1. Lyric2 will
always remain at the same volume setting unless
changed as described.
Switch OFF
To switch the device OFF, switch to SLEEP-mode (2 beeps)
and continue holding the magnet close to the device.
After a pause of 2-3 seconds the device will switch OFF
(4 beeps). To switch the device back ON, hold the magnet
briefly and then remove it (1 beep).
Example
Lyric2 is switched ON, VOL set to 2, MaxVol set to 4,
patient wants to change volume from level 2 to 3.
Change to SLEEP-mode (2 beeps), switch device back
ON (1 beep) and keep holding the magnet close to the
device. After a 2-3 second break, Lyric2 starts to cycle
through the volume levels. Count the beeps and remove
magnet after the third beep which indicates level 3.
If third beep was missed, continue to hold the magnet
close to the device. The device will go up to level 4
and then restart the cycle at level 1. Count the beeps
again and remove the magnet when level 3 is reached.
Remove Lyric2
See chapter "Counseling instruction" on the next page
for the patient instruction on how to remove Lyric2 from
from the ear canal using the SoundLync.