Acticare TSE User Manual page 37

Table of Contents

Advertisement

Standard TSE and TENS Therapy Mode Parameters
Mode
Waveform
Frequency
FRQ
TSE1
2.5kHz
TSE2
10kHz
TSE3
250Hz
TENS1
100Hz
TENS2
5Hz
TENS3
100Hz
TENS4
100Hz
TENS5
100Hz
TENS6
100Hz
HFT1
500Hz
HFT2
500Hz
See page 26 for diagrams of Waveform shapes and Burst parameters. Modulation modes
produce a linear variation of pulse width between 10 and 50µS over the specified period.
TSE and TENS modes - Peak Output into 500ohms
Pulse width
Peak Voltage
0.5 - 2µS
250V
3 - 5 S
µ
200V
6 - 10µS
160V
IC Mode Program Presets
Programs Available (AMF= Amplitude Modulation Frequency)
Name
Carrier
PROG1
4kHz
PROG2
4kHz
PROG3
4kHz
PROG4
4kHz
PROG5
4kHz
PROG6
4kHz
PROG7
2kHz
PROG8
2kHz
VAR
2kHz or 4kHz
Waveform
Pulse
Type
Width
WAVE
PULS
Biphasic
1.5 S
Biphasic
0.5 S
Biphasic
4 S
Biphasic
50 S
Biphasic
200 S
Biphasic
100 S
Biphasic
100 S
Biphasic
10-50 S
Biphasic
10-50 S
Biphasic
20 S
Biphasic
20 S
Pulse width
20 S
50 S
100 - 200µS
Base AMF
0
0
0
10
50
100
0
0
0Hz to 250Hz
34
Technical Information
Interpulse
Burst
Spacing
Time
SPAC
BT
200 S
OFF
0.5 S
OFF
2000 S
OFF
0 S
OFF
0 S
OFF
0 S
100mS
0 S
500mS
0 S
3 sec modulation
0 S
6 sec modulation
0 S
OFF
0 S
50mS
Peak Voltage
130V
90V
70V
Peak AMF
Pattern
10
6^6
25
6^6
100
6^6
150
6^6
200
6^6
150
6^6
10
6^6
25
6^6
0Hz to 250Hz
1/1, 6/6, or 6^6
Burst
Freq
BF
5Hz
1Hz
10Hz
12

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tens

Table of Contents