Changing Therapy Modes
1. Check the device is in IC Mode
on page 11). Turn the device on by pressing
and releasing the on/off button
2. Enter the Therapy Mode Menu
the
and TIMER
same time
. The display flashes "MODE
MENU" as confirmation.
3. Press
or
to select the desired
program
from the table opposite. If you
choose a preset program press and hold
TIMER
to confirm your selection and exit
the Therapy Mode menu.
4. If you choose the variable user
programmed mode (VAR)
set the treatment parameters in the following
order: Carrier frequency, Base Amplitude
Modulated Frequency (AMF), Peak AMF,
Sweep pattern. See the diagram above right.
Scroll through parameters using the TIMER
key and alter the parameter value using the
or
keys.
Pressing the TIMER
cycle through the parameters so you can
make any changes or check them.
5. Save your VAR program parameters and
exit Therapy Mode
menu by pressing and
holding the TIMER
new settings press
and TIMER
You can check the settings by swithcing the
device off and on using
(see STEP 1
.
by pressing
buttons
together at the
If you see the timer
screen instead of the
mode menu, you did
not press
&
together. Press
to exit & try again.
you will need to
key continuously will
button. To cancel the
together.
.
Interferential Mode
Sweep Pattern Options
6/6
Peak
Base
0
1
1/1
Peak
Base
0
6^6
Peak
Base
0
For a constant sine wave output at
the carrier frequency, set peak and
base at 0Hz in VAR mode.
Variable (VAR) Mode Menus
Example Screen
10
5
6
11
12
17
18
Time (seconds)
1
2
3
Time (seconds)
6
12
18
Time (seconds)
Parameter
Carrier Frequency
2kHz or 4kHz
Base Amplitude
Modulated
Frequency (AMF)
0Hz - 250Hz
Peak Amplitude
Modulated
Frequency (AMF)
0Hz - 250Hz
Sweep Pattern
6^6, 6/6, or 1/1