Technical Information; General Specifications - Acticare TSE User Manual

Table of Contents

Advertisement

Technical Information

IC modes
Peak current
Waveform type
Carrier frequencies
AMF range
Sweep patterns
Treatment Timer
TSE and TENS modes
Average current limit
Maximum charge per pulse
Waveform type
Treatment Timer
Pulse width
Output frequency
Safety
Device Classification
Key Design Compliance
Electrical protection rating
Environmental
Temperature
Relative humidity (non-condensing)
Atmospheric pressure

General Specifications

12
Internal power source
Cradle input voltage
Europe / UK / Australian Versions
North American Version
Cradle output voltage
Dimensions / Weight
All electrical specifications are +/- 10% into a 500 ohm load
100mA into 500ohm load @ 65V
Sinusoidal
2kHz and 4kHz
0 to 250Hz in 5Hz steps
6^6, 6/6, 1/1
1 to 30 min
20mA @ 250VDC bus
25µC per pulse under normal operation
75µC with single component failure
Biphasic, monophasic or twin peak
square wave
10 mins to 3 hours
0.5µS to 200µS
1Hz to 500kHz
IIa Medical Devices Directive (93/42/EEC)
EN 60601-1:1990, EN 60601-2-10
ISO EN ISO 14971:2000
Type BF equipment (EN 60601-1).
Not for connection to mains power
except through approved cradle.
Operating
Shipping and Storage
10C to 50C
-40C to 70C
(Note: maximum 40C when charging)
30-75%
10-100%
860-1060 hPa
500-1060 hPa
Four 1.2V-1.5V AA batteries >1500mA
capacity (rechargeable or alkaline)
240-250VAC 50-60Hz
115VAC 60Hz
6V @ 1000mA
100x72x38mm / 235gms (device only)
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tens