General Functions - KERN PLS Operating Instructions Manual

Pls; plj; als; alj electronic precision and analytical balance
Hide thumbs Also See for PLS:
Table of Contents

Advertisement

8.2 General functions

Menu selection:
Turn on balance using the ON/OFF-button
Press the Tare-button; the display shows horizontal lines
While this display is on, press the Cal-button
Keep F-button pressed until the desired selection is displayed
Function
Rear illuminated
Display
(see Chapter. 7.1.1)
Auto-zero-function
(see Chapter 6.5.4)
Setting the automatic
adjustment function,
only ALJ /PLJ6100-2
in non-verified setting
(see Chapter 6.7.1)
Data output
(only ALJ / PLJ6100-2
in non-verified setting)
Vibration filter
Forming of an
average value
(Animal weighing
program)
Display speed
Standard weighing
unit
* = Factory setting
° = Recommended counter setting
62
Display
Selection
PRINT-button
F-button
F-button
0
bl
1
AUTO 0*
AUTO
AUTO 1
Ac_t 0*
Ac_t
Ac_t 1
AC_c 0*
Ac_c
AC_c 1
Pd_d 0*
Pd_d
Pd_d 1
Con 1
Con 2
Con 3*
Con
Con 4
Con 5°
AuE 1
AuE 2*
AuE
AuE 3
FIL 0*
FIL
FIL 1
St_uni
Description of the
selection options
off
on
Auto-zero-
function activated
Auto-zero-
function deactivated
Automatic adjustment function
activated
Automatic adjustment function
deactivated
Automatic adjustment function
activated
Automatic adjustment function
deactivated
Last digit is not printed
Last digit is printed
Sensitive and fast
(very quiet set-up location)
Robust but slow (very busy
set-up location)
Fastest display
Slowest display
Setting for dosing
deactivated
Setting for dosing
activated
see Chapter 7.2.2
PLS/PLJ/ALS/ALJ-BA-e-0819

Hide quick links:

Advertisement

Table of Contents
loading