6.5.3 Balance zero-display

If the balance does not show exactly zero although the pan scale is unloaded, press
the TARE-button and the balance will be reset to zero [ a ].

6.5.4 Auto-zero-function

In the menu you can turn on or off the auto-zero-function. To do so, proceed as
follows:
Turn on balance using the ON/OFF-button
Press the Tare-button; the display shows horizontal lines
While this display shows, press the CAL-button, until AUtO is displayed
Press the PRINT-button; the display now shows the current menu item
blinking
Use the F-button to choose between the following settings:
AUtO 0: Auto-zero-function activated
AUtO 1: Auto-zero-function deactivated
Use the PRINT-button to confirm the selected setting

6.6 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. It is
also recommendable to adjust the balance periodically during weighing operation in
order to obtain exact measured values.