1. INSTALLATION INSTRUCTION 1.3.1 This scale is powered by the built-in rechargeable battery and/ or by a AC CHECK AND MAKE SURE THA T THE FOLLOWING adapter included. ITEMS ARE INCLUDED. 1.3.2 Chec k and make sure that the voltage of This operation manual the AC adaptor supplied matches with the Scale x 1...
4. KEYBOARD LAYOUT KEYS Press this key to turn scale off. ON/ZERO Press this key to turn scale on or set weight to zero (max. zero range = 2% of rated capacity) or decrease value during function setting. TARE Press this key to tare off the weight of a container or increase value during function setting.
Page 5
various function value stable. setting. 12 HI INDICATOR This sign will on when DISPLAY SIGNSZERO STATUS SIGN. When the weight detected exceeds the à0ß weigh detected is at true upper weight limit. zero, an arrow will appear and point at this sign. 13 GO INDICATOR This sign will on when weight detected falls within MINUS SIGN.
Output Port printer interface. INTERNAL FUNCTION TABLE Function Symbol Description Note Fun - 1 Scale configuration setting dealer calibration. Fun – 2 Display segment check. Fun – 3 Off set value and span value reading. Fun – 4 Auto power setting.
Page 7
Function when RS232 interface is purchased. e. Press TARE until required Function when printer interface setting/value appears. purchased. f. Press SET to confirm. h. Repeat step c to step f for other function 6.2 HOW TO ENTER THE REQUIRED FUNCTION MODE setting, or 6.2.1 Function 1 to 6 j.
Page 8
Fun-5 Set RS232 baud rate: 7 rates are Under this mode, no data will be available: transmitted if weight detected 300, 600, 1200, 2400, 4800, 9600 and is less than 40 d. 19200 02 = Continuos transmission mode. Under this mode, data Fun-6 Contact...
Refer to Appendix A for minimum Beep 1: Alarm when weight detected weight required. falls within the upper and lower NOTE : Details every transaction limit. transmitted will be stored in memory. Beep 2: Alarm when weight detected Press MODE to transmit the total value exceeds the upper limit or below accumulated to printer.
Page 10
When backlight is on, press SET to turn For best weighing result, always place backlight off. this scale on a level surface. This scale TARE FUNCTION is equipped with a bubble level. If 7.4.1 To Tare Off the Weight of a Container required, adjust the adjustable feet underneath the scale to obtain a level Place a container onto platter, then...
Page 11
or overloading the scale may cause obtained. The unit piece weight forms damage to the weight sensor inside. the foundation of quantity calculation. It is a good practice to remove all load a. If a container is used, place the from platter when not in used.
Page 12
information. place more/remove some subject matter onto the platter. The new quantity will 7.8.3 Counting Transact ion be shown on the display. Put more subject matter or remove part of the subject matter, the new piece 7.9.1.2 The Auto Unit Piece Weight Enhancement value is shown on the display.
Page 13
turned scale off will not result in the 7.10.1 Set Audio Alarm Output lost of weight limits set. If an audio alarm output is required, NOTE: Weight Check will not function if a enter function 30 and set appropriate weight unit, other than the weight unit parameter.
Page 14
Sampling is the first step of Percentage subject matter, relative Check function. By means of sampling percentage value will appear on the process, the weight of subject matter display. at 100% is determined. Follow the below NOTE: For best checking result, refer to steps for sampling.
Page 15
20 LINE INPUT #1, A$ INTERNAL FUNCTIONS for more 30 PRINT A& information. 7.13.2Connecting the Scale with a Computer 40 GOTO 20 50 END a. Turn scale off b. Save the above program file. b. Turn printer off c. Connect the printer output of scale to 7.12.3 Connecting the Scale with a Computer printer by a appropriate data cable Follow the below steps to connect the...
recharged. b. GREEN: Rechargeable battery is completely charged. Check Is a right AC adaptor employed or is it inserted properly both to the wall outlet and into the scale properly? Action Check AC adaptor and secure both ends of the AC adaptor into the wall outlet and DC input of scale.
Page 17
Check Action Is the scale overloaded? Contact your dealer. Action Remove all load on platter and SCALE DOES NOT COUNT CORRECTLY check again. Check Check Is the platter insert correctly Is the right sample size employed into the scale? during sampling process? Action Action Take out the platter and insert...
Check Is the LOW BATTERY sign on? 9.2 Do not lift/move scale by holding the Action Recharge the battery or employed platter. the AC adaptor. 9.3 Do not submerge scale into water. DATA TRANSMISSION ERROR 9.4 Take platter away from scale before Check Are all transmission parameters transportation.
Page 19
Model No. Minimum Weight Required for RS=Not stable Mode 00 Mode 01 GW-1500 GW-3000 = Polarity of value (Plus or Minus) GW-7500 + or blank = positive value GW-15K GW-30K = Negative value Appendix B RS232 Data Format of Function 20 =Value of weight, pieces or percentage 12 13 =Unit...
Page 20
Unit (Note A) Function 0001. 10.00 0002. 20.00 (Note B) Model No. Minimum Weight 0003. 30.00 Required 0003/ 60.00 (Note C) Gw-1500 GW-3000 GW-7500 GW-15K NOTE A: S/N = Sequence Number GW-30K Unit = The current function being employed e.g, k.g, g, lb, pc, or Appendix E Feather Weight of Counting and Percentage Check NOTE B: Weig ht check result:...