Membrane Output Calculation Example
What is the expected GPD from a 75 GPD System at 40 psi pressure and 60°F
water temperature?
PCF = 40 ÷ 60 = 0.666
TCF = 0.754 (from Table 1)
Expected GPD = 75 × 0.666 × 0.754 = 37.7 GPD ± 15%
37.7 GPD would be the Actual Production Rate
Performance Test
The performance of a RO membrane is measured by its ability to reject salts
(or TDS (Total Dissolved Solids)).
Important: Test the quality of the membrane once every 6 months.
Procedure:
1.
Measure tap water TDS. (Call it X)
2.
Run the system for 15-20 minutes.
3.
Measure RO water's TDS. (Call it Y).
4.
Subtract RO water TDS from tap water TDS. (X - Y)
5.
Divide this quantity by tap water TDS. (X - Y) ÷ X
6.
Rejection = [(X - Y) ÷ X ] ×100
Rejection of the RO Membrane Calculation Example
1.
Tap water TDS = 150 ppm (X)
2.
RO water TDS = 7 ppm (Y)
3.
X - Y = 143 ppm
4.
(X - Y) ÷ X = 143 ÷ 150 = 0.953
5.
Rejection = [ ( X - Y) ÷ X ] ×100 = 0.953 ×100 = 95.3
Membrane TDS Rejection = 95.3 % : Rejection rates less than 95% may
indicate that the membrane should be replaced.
* IF YOU NEED HELP WITH THIS PROCEEDURE, PLEASE CONTACT SPECTRAPURE*
SpectraPure®Inc.
2167 East Fifth St, Tempe, Arizona 85281
20
®
480.894.5437 Call us toll-free 1.800.685.2783

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROPLUS RO System and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Proplus di systemProplus ro/di system

Table of Contents