TROUBLESHOOTING & FAULT REPAIR
P&G S-Drive 45 Amp controller
the operating conditions of your scooter. If it detects a problem it will indicate with error message by flashing
light on the power ON/ OFF light. You must count the number of the flash, and see the list to check what kind
of error has happened according to the number)
Number of
Fault
Flashes
1
Battery needs
recharging
2
Battery
voltage too
low
3
Battery
voltage too
high
4
Current limit
time out
5
Brake fault
: You r scooter is fitted with S-Drive controller, which continuously monitors
Impact
on
Scooter
Will drive
Drive
inhibited
Drive
inhibited
Drive
inhibited
Drive
inhibited
22
Notes
Battery charge is running low. Recharge the batteries
as soon as possible.
Battery charge is empty. Recharge the batteries. If the
scooter is left off for a few minutes, battery charge
may recover sufficiently to allow driving for a short
period of time.
Battery charge is too high. If a charger is plugged in,
unplug it or turn the Charge/Run switch to Run.
Scooters powered by RHINO will charge the batteries
when traveling down slopes or decelerating.
Excessive charging in this manner may cause this
fault. Turn the scooter power off and then back on
again.
The scooter has drawn too much current for too long,
possibly because the motor has been over worked,
jammed or stalled. Turn the scooter power off, leave
for a few minutes, and then turn the power back on
again.
The controller has detected a shorted motor. Check
the loom for shorts and check the motor. Contact your
service agent.
Check that the park brake release lever is in the
engaged position.
The park brake coil or wiring is faulty. Check the
park brake and wiring for open or short circuits.
Contact your service agent.