Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the HP-IB Address
To configure the power supply for the HP-IB interface, proceed as follows:
I/O
Config
1 Turn on the remote configuration mode.
HP-IB / 488
You will see the above message on the front-panel display if the power supply has not
been changed from the factory setting. If "RS-232" appears, choose "HP-IB / 488"
by turning the knob to the right.
I/O
Config
2 Move to the HP-IB address setting mode.
ADDR 05
The address is set to "05" when the power supply is shipped from the factory. Notice
that a different HP-IB address may appear if the power supply has been changed from
the factory setting.
3 Turn the knob to change the HP-IB address.
The displayed address is changed when turning the knob to the right or left.
I/O
Config
4 Save the change and turn off the I/O configuration mode.
CHANGE SAVED
The address is stored in non-volatile memory, and does not change when power has
been off or after a remote interface reset. The power supply displays a message to
show that the change is now in effect. If the HP-IB address is not changed, "NO
CHANGE" will be displayed for one second.
N o t e
To exit the I/O configuration mode without any further changes, press the "I/O
Config" key until the "NO CHANGE" message is displayed.