Remote Voltage Sensing at the Front and Rear Terminals
3

Remote Voltage Sensing at the Rear Panel

External sense terminals are also available on the back of the power supply that allow
the rear output voltages to be sensed at the load, which compensates for impedance
losses in the load wiring. The front panel binding posts are paralleled with the rear
output terminals.
The rear output terminals accept wires sizes from AWG 22 to AWG 10.
To minimize the possibility of instability on the output, keep load leads as short as
possible and bundle or twist the leads tightly together to minimize inductance.
Figure 3-3. Rear local sensing connections
N o t e
For rear local voltage sensing connections, the front shorting bars must be removed
first and connect the sense wires as shown in Figure 3-3.