Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
Programming Overcurrent Protection
Overcurrent protection guards the load against output currents that reach a specified
value greater than the programmed protection level. It is accomplished by
programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how to
check OCP operation and how to clear overcurrent condition.
• Front-panel operation:
Setting the OCP Level and Enable the OCP Circuit
Power
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF
** annunciator turns on); and the knob is selected for voltage control.
25V
Output
2 Enable the output.
On/Off
The
Over
3 Enter the OCP menu and set the trip level.
Current
LEVEL 22.0 A
LEVEL 7.5 A
You will see the above message on the display when you enter the OCP menu. Adjust
the knob for the desired OCP trip level.
Over
4 Enable the OCP circuit.
Current
OCP ON
You will see the above message after pressing the
*For HP E3633A Model
34
annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
annunciator turns off and the display will go to the meter mode.
OFF
(E3633A)
(E3634A)
**For HP E3634A Model
8V
Over
key.
Current
* or