Chapter 5 Error Messages
Execution Errors
-141
Invalid character data
Either the character data element contained an invalid character or the particular
element received was not valid for the header.
-144
Character data too long
The character data element contained too many characters.
-148
Character data not allowed
A discrete parameter was received but a character string or a numeric parameter was
expected. Check the list of parameters to verify that you have used a valid parameter
type.
Example: DISP:TEXT ON
-151
Invalid string data
An invalid character string was received. Check to see if you have enclosed the
character string in single or double quotes.
Example: DISP:TEXT 'ON
-158
String data not allowed
A character string was received but is not allowed for the command. Check the list
of parameters to verify that you have used a valid parameter type.
Example: TRIG:DEL 'zero'
-160 to -168
Block data errors
The power supply does not accept block data.
-170 to -178
Expression errors
The power supply does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored.
Make sure that the trigger source should be selected to the bus and the trigger
subsystem should be initiated by INIT[:IMM] command.
-213
Init ignored
An INITiate command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a measurement in
progress and place the power supply in the ''idle'' state.