Chapter 4 Remote Interface Reference
Calibration Commands
CALibration:VOLTage:LEVel {MINimum | MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output state is
ON. It sets the power supply to a calibration point that is entered with CAL:VOLT
command. During calibration, three points must be entered and the low-end point
(MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of the power supply. It
takes about 10 seconds to execute the command. The calibration must be unsecured
and the output be opened before calibrating the overvoltage protection circuit. The
power supply automatically performs the calibration and stores the new overvoltage
constant in nonvolatile memory. Notice that voltage calibration precedes before
sending this command.