Chapter 3 Front-Panel Operation

Disabling the Output

Disabling the Output

The output of the power supply can be disabled or enabled from the front panel.
When the power supply is in the "Off" state, the
output is disabled. The
OFF
the "On" state. When the output is disabled, the voltage value is 0 volts and the current
value is 0.02 amps.
The output state is stored in volatile memory; the output is always disabled when
power has been off or after a remote interface reset.
While the output is disabled, the range selection keys, the control knob, resolution
selection keys, and adjust selection key are still working. If the display is in the
meter mode, you cannot see the changes of output voltage and current settings on
the display when turning the knob. To see or check the changes when the outputs
are disabled, the display should be in the limit mode.
• Front-panel operation:
You can disable the output by pressing
"Off" and "On" states.
• Remote interface operation:
OUTP {OFF|ON}
40
annunciator turns on and the
OFF
annunciator turns off when the power supply returns to
Output
key. This key toggles between output
On/Off
Disable or enable the output