Volt:prot - Agilent Technologies E4356A Operating & Programming Manual

Telecommunications dc power supply
Table of Contents

Advertisement

MAX
Returned Parameters

VOLT:PROT

This command sets the overvoltage protection (OVP) level of the power supply. If the output voltage exceeds the OVP
level, then the power supply output is disabled and the Questionable Condition status register OV bit is set (see "Chapter 8 -
Status Reporting"). An overvoltage condition can be cleared with the OUTP:PROT:CLE command after the condition
that caused the OVP trip is removed. The OVP always trips with zero delay and is unaffected by the OUTP:PROT:DEL
command.
Returned Parameters
84
Language Dictionary
Command Syntax
Parameters
Default Suffix
*RST Value
Examples
Query Syntax
Related Commands
Command Syntax
*Alternate Syntax
Parameters
Default Suffix
*RST Value
Examples
Query Syntax
Related Commands
* Available to accommodate earlier power supply programs.
[SOURce]:VOLTage[:LEVel][:IMMediate][AMPLitude] <NRf+>
[SOURce][:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
Table 7-1
V
Table 7-1
VOLT 200 MA
VOLTAGE:LEVEL 200 MA
VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5
VOLT:TRIG 20
VOLTAGE:LEVEL:TRIGGERED 20
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? MIN
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]?
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]? MAX
[SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]? MIN
<NR3> VOLT? and VOLT:TRIG? return presently programmed
immediate and triggered levels. If not triggered level is programmed,
both returned values are the same.
VOLT? MAX and VOLT? MIN return the maximum and
minimum programmable immediate voltage levels.
VOLT:TRIG? MAX and VOLT:TRIG? MIN return the
maximum and minimum programmable triggered voltage levels.
For VOLT
*SAV
*RCL
For VOLT:TRIG
ABOR
[SOURce]:VOLTage:PROTection[:LEVel] <NRf+>
[SOURce]:VOLTage:PROTection:AMPLitude <NRf+>
Table 7-1
V
MAX
VOLT:PROT 21.5
VOLT:PROT:LEV MAX
VOLTAGE:PROTECTION:LEVEL 145E-1
[SOURce]:VOLTage:PROTection[:LEVel]?
[SOURce]:VOLTage:PROTection [:LEVel]? MIN
[SOURce]:VOLTage:PROTection [:LEVel]? MAX
<NR3> VOLT:PROT? returns presently programmed OVP level.
VOLT:PROT? MAX and VOLT:PROT? MIN return the
maximum and minimum programmable OVP levels.
OUTP:PROT:CLE
*RST
*RST
VOLT
*RST
*SAV
*RCL

Advertisement

Table of Contents
loading

Table of Contents