Initiate Subsystem; Init Init:cont; Measure Subsystem; Meas:curr? Meas:volt - Agilent Technologies E4356A Operating & Programming Manual

Telecommunications dc power supply
Table of Contents

Advertisement

Initiate Subsystem

This subsystem enables the trigger system.
INIT
INIT:CONT
When a trigger is enabled with this command, an event on a selected trigger source causes the specified trigging action to
occur. If the trigger subsystem is not enabled, all trigger commands are ignored. If INIT:CONT is OFF, then INIT
enables the trigger subsystem only for a single trigger action. The subsystem must be enabled prior to each subsequent
trigger action. If INIT:CONT is ON, then the trigger subsystem is continuously enabled and INIT is redundant.

Measure Subsystem

This subsystem returns output measurements.
MEAS:CURR?
MEAS:VOLT?
These queries return the voltage and current measured at the power supply's sense terminals.
Returned Parameters
76
Language Dictionary
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
Query Syntax
Parameters
Default Suffix
Examples
MEAS:CURR?
<NR3>
INITiate[:IMMediate]
INITiate:CONTinuous <bool>
For INIT[:IMM]
(None)
For INIT:CONT 0|1|OFF|ON
OFF
INIT INITIATE:IMMEDIATE
INIT:CONT 1
INITIATE:CONTINUOUS 1
For INIT[:IMM]
(None)
For INIT:CONT
INIT:CONT?
<NR1>
0|1
ABOR <GET> *RST
MEASure:CURRent[:DC]? MEASure:VOLTage[:DC]?
(None)
A for MEAS:CURR?
MEAS:VOLT?
TRIG
*TRG
V for MEAS:VOLT?
MEASURE:VOLTAGE:DC? MV