Specifications - Agilent Technologies E4356A Operating & Programming Manual

Telecommunications dc power supply
Table of Contents

Advertisement

Specifications

Specifications are performance parameters warranted over the specified temperature range.
Supplemental Characteristics are not warranted but are descriptions of performance determined either by design or type
testing.
Parameter
Output Ratings
(@ 0 to 45°C)
Programming Accuracy
(@ 25 °C ± 5 °C)
Ripple & Noise
(from 20 Hz to 20 MHz with outputs ungrounded,
or with either output terminal grounded)
Readback Accuracy
(from front panel or over GPIB with respect to
actual output @ 25 °C ± 5 °C)
Load Regulation
(change in output voltage or current for
any load change within ratings)
Line Regulation
(change in output voltage or current for
any line change within ratings)
Transient Response Time
(for the output voltage to recover to its previous level within 0.1% of the rated voltage or
20 mV, whichever is greater, following any step change in load current up to 50% of the
rated current)
Table A-1. Performance Specifications for Agilent E4356A
Voltage:
Current:
Voltage:
Current:
Constant Voltage rms:
Constant Voltage p-p:
Constant Current rms:
(measured with 60 cm leads )
Voltage:
Current:
Voltage:
Current:
Voltage:
Current:
A
Value
0 - 80 V (0 to 26A)
0 - 30 A (0 to 70 V)
0.04% + 80 mV
0.l % + 25 mA
2 mV
16 mV
25 mA
0.05% + 120 mV
0.l % + 35 mA
0.002% + 3 mV
0.005% + 2 mA
0.002% + 3 mV
0.005% + 2 mA
< 900 µs
Specifications 93