*IDN?

Meaning and Type
Identification Query System Interface
Description
This query requests the power supply to identify itself. It returns a string composed of four fields separated by commas.
Returned Parameters

*OPC

Meaning and Type
Operation Complete Device Status
Description
This command causes the interface to set the OPC bit (bit 0) of the Standard Event Status register when the power supply
has completed all pending operations. (See *ESE for the bit configuration of the Standard Event Status register.) Pending
operations are complete when:
All commands sent before *OPC have been executed. This includes overlapped commands. Most commands
are sequential and are completed before the next command is executed. Overlapped commands are executed in
parallel with other commands. Commands that affect output voltage, current or state, relays, and trigger actions are
overlapped with subsequent commands sent to the power supply. The *OPC command provides notification that
all overlapped commands have been completed.
Any change in the output level caused by previous commands has been completed (completion of settling
time, relay bounce, etc.)
All triggered actions are completed
*OPC does not prevent processing of subsequent commands but Bit 0 will not be set until all pending operations are
completed.
Related Commands
66
Language Dictionary
Query Syntax
Example
Related Commands
Command Syntax
*OPC
Parameters
(None)
*OPC? *WAI
*IDN?
<AARD>
Field
Information
Agilent Technologies Manufacturer
xxxxA
4-digit model number followed by a
letter suffix
nnnnA-nnnnn
10-character serial number or 0
<R>.xx.xx
Revision levels of firmware
Agilent Technologies,6681,0,A.00.01
(None)