Agilent Technologies E4356A Operating & Programming Manual page 120

Telecommunications dc power supply
Table of Contents

Advertisement

1
ARPS Command
VSET x
VSET xV
VSET xMV
ISET x
ISET xA
ISET xMA
VSET?
ISET?
VOLT?
IOUT?
OVP x
OVP xV
OVP xMV
OVP?
VMAX x
VMAX xV
VMAX
xMV
IMAX x
IMAX xA
IMAX
xMA
VMAX?
IMAX?
DLY x
DLY xS
DLY xMS
DLY?
OUT OFF
OUT 0
OUT ON
OUT 1
FOLD OFF
FOLD CC
FOLD 2
FOLD 0
FOLD CV
FOLD 1
FOLD CC
FOLD 2
FOLD?
1
x = any digit (within range) MA = milliampere MV = millivolt MS = millisecond.
120 Compatibility Language
Table G-1. ARPS Commands
These commands program output voltage. See Table 7-1 for the
programming ranges of these commands.
Initial condition: 0 V
These commands program output current. See Table 7-1 for the
programming ranges for these commands.
Initial condition: 0 A
These commands read voltage or current settings.
These commands measure and read output voltage or current.
NOTE: OVP commands do not work with Agilent 603xA
supplies. These commands program the overvoltage protection.
The OVP setting is programmed in either volts or millivolts. See
Table 7-1 for the programming ranges of these commands.
Initial condition: 10% above rated output.
This command reads the OVP setting.
These commands program an upper limit (soft limit) to the voltage
programming value that the power supply will accept. The
programming ranges are the same as those used for VSET.
These commands program an upper limit (soft limit) to the current
programming value that the power supply will accept. The
programming ranges are the same as those used for ISET.
These commands read the soft voltage or current limits.
These commands program the delay time before a new output
voltage or current is implemented or an RST, OUT ON, or CLR
command is received. During the delay the CV, CC, and CR
conditions cannot be reported as faults, and foldback protection is
disabled.
This command reads the delay time setting.
These commands enable or disable the power supply output. The
disabled state programs the output to relatively low voltage and
current values. The unit will be able to implement commands even
while the output is disabled.
Initial condition: OUT ON
These commands enable or disable Foldback protection.
Foldback protection disables the power supply output if the power
supply switches to whichever mode (CV or CC) is defined as the
fold (error) condition. Note that foldback protection is disabled
during the DELAY period.
Initial condition: FOLD OFF
This command reads the Foldback setting.
Description
Similar SCPI
Command
VOLT
CURR
VOLT?
CURR?
MEAS:VOLT?
MEAS:CURR?
VOLT:PROT
VOLT:PROT?
(None)
(None)
(None)
OUTP:PROT:DEL
OUTP:PROT:DEL?
OUTP:STAT OFF
OUTP:STAT 0
OUTP:STAT ON
OUTP:STAT 1
(None)
(None)