8. Connect A battery harness to the new rear battery according to the battery wiring diagram. See figure 31.
WARNING! Make sure you tighten the fasteners so that the connections are secure.
9. Install the new batteries.
10. Reinstall the front cover and foot platform.
11. Charge the batteries. See VI. "Batteries and Charging."
When to See Your Quantum Rehab Dealer for Service
The following symptoms could indicate a serious problem with your power chair. If necessary, contact your Quantum
Rehab Dealer. When calling, have the model number, serial number, nature of the problem and the error code if available.
Motor noise
Frayed harnesses
Cracked or broken connectors
Uneven wear on any of the tyres
Jerky motion
Pulling to one side
Bent or broken wheel assemblies
Does not power up
Powers up, but does not move
Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:
Check the harness connections. Make sure they are tight.
Check the circuit breaker. Reset it if necessary.
Check the battery connections.
If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are available
at automotive parts stores. Disconnect both batteries before load testing and follow the directions that come with the load
tester. If either one of the batteries fails the load test, replace both of them. If your power chair still does not power up,
contact your Quantum Rehab Dealer.
Quantum 6000 Series
C A R E
A N D
M A I N T E N A N C E
www.pridemobility.com