Original – Original signal is used to generate the B-Scan.
Envelope – Envelope signals is used to generate the B-Scan.
SAFT
When activated, applies a Synthetic Aperture Focusing Technique to the
raw data to produce a sharper image.
SAFT uses path length and positioning information to correct the image.
The quality of the final image is dependent on the spacing of the meas-
urements.
© 2014 Proceq SA
3.4

Measurement Screen

The standard measurement screen is shown on page 3. All settings
are directly accessible from the measurement screen.
Zoom
Zoom in by placing thumb and index finger together on the
screen and spreading them apart. This can be used in both the
horizontal and vertical directions when making a measurement.
Zoom out by placing thumb and index finger apart on the
screen and pinching them together.
Pan
Pan the image from left to right by dragging.
Measuring screen controls (see page 3)
1
Filename: Enter the file name and press return. Saved measure-
ments will be stored with this file name. If several measurements are
made under the same file name, a suffix increments after each measure-
ment.
2
Measurement mode: Select the type of measurement to be carried
out (see section "3.5 Measurement Modes").
3
The top right hand corner of the display shows the current trans-
ducer selected, current time and the battery status.
4
Gain: Adjust the receiver gain, from 1x up to a maximum of 10 000x.
5
Voltage: Adjust the transmitter voltage. For best results, it is best to
begin with low transmitter voltage and a low gain setting. Then increase
until a stable signal level is achieved. Signal clipping should be avoided.