Options; Table 4. Typical Deepest Operating Temperature - Princeton PIXIS-XB System User Manual

Table of Contents

Advertisement

Appendix A
Power (DIN): 12 VDC (6.6A max) input from power supply for small array cameras
Power (3-pin LEMO): 12 VDC (12.5A max) input from power supply for large
Fan: (Air-cooled systems only) 24 CFM fan capacity at full power.
Coolant Ports: (Liquid-cooled systems only) Two interchangeable
quick disconnect fittings
Deepest Operating Temperature (with ambient air at +20°C):
Temperature Stability: ±0.05°C; closed-loop stabilized-temperature control
Power Input: 100-240 VAC; 47 to 63 Hz, 1.9A. DC power to camera is provided by the
self-switching power supply.
Tripod Mount: 1/4-20 x .25" mounting hole at bottom of camera. M6 threaded adapter
supplied with system.
A/D Converters: Dual digitizers with 100 kHz/2 MHz readout rates. Software-
selectable. Low-speed operation gives better noise performance; high-speed
operation allows faster data acquisition.

Options

Contact the factory for information regarding options available for your system.
Basic Specifications
(PIXIS-XB:400 or PIXIS-XB: 1024).
array cameras (PIXIS-XB:1300 or PIXIS-XB:2048).
Pin 1: 12 VDC
Pin 2: Ground
Pin 3: Chassis
for connection to the CoolCUBE
CCD Size
1024x1024
1340 x 400
1340 x 1300
2048 x 2048

Table 4. Typical Deepest Operating Temperature

Pin 3, 5: 12 VDC
Pin 1, 2, 4: Return
Shell: Ground
Typical
-90°C
-90°C
-65°C (air-cooled)
-70°C (CoolCUBE
)
II
-65°C (air-cooled)
-70°C (CoolCUBE
)
II
1/4" CPC valved
circulator hoses.
II