Acquisition Started But Viewer Contents Do Not Update; Baseline Signal Suddenly Changes; Camera Stops Working; Figure 44. Acquisition Display - Princeton PIXIS-XB System User Manual

Table of Contents

Advertisement

64

Acquisition Started but Viewer Contents Do Not Update

In LightField, live data is normally displayed in the Experiment workspace viewer as it is
being acquired (Preview or Acquire mode). If the viewer is not being updated and
acquisition is occurring, check to see if there is a filename in the top row of the viewer: in
the figure below, the filename 2010 May 11 13_13_49.raw is displayed. If there is a
filename, click on the camera icon. The file data will be cleared from the viewer and the
live data will then be displayed.

Baseline Signal Suddenly Changes

A change in the baseline signal is normal if the temperature, gain, or speed setting has
been changed. If this occurs when none of these settings have been changed, there may
be excessive humidity in the camera vacuum enclosure. Turn off the camera and contact
Princeton Instruments Customer Support. See page 92 for contact information.

Camera Stops Working

Problems with the host computer system or software may have side effects that appear to
be hardware problems. If you are sure the problem is in the camera system hardware,
begin with these simple checks:
If the system still does not respond, contact Princeton Instruments Customer Support.
PIXIS-XB System Manual
Turn off all AC power.
Verify that all cables are securely fastened.
Turn the system on.

Figure 44. Acquisition Display

Version 2