Controller Is Not Responding; Coolcube Ii : Low Coolant (Air In The Hoses) - Princeton PIXIS-XB System User Manual

Table of Contents

Advertisement

66

Controller Is Not Responding

If this message pops up when you click on OK after selecting the "Interface Type" during
Hardware Setup (under the WinView Setup menu), the system has not been able to
communicate with the camera. Check to see if camera has been turned ON and if the
USB 2.0 interface card, its driver, and the USB cable have been installed.
CoolCUBE
: Low Coolant (Air in the Hoses)
II
WARNING!
If more than two inches (50.8 mm) of the coolant line is filled with air, the pump will
stop working and may be damaged. If flow stops while the pump is on, turn off the
CoolCUBE
1. Unscrew the reservoir cap (on top of the CoolCUBE
2. Screw the reservoir cap back in.
3. Make sure the power switch is turned off before plugging the circulator in.
4. Plug the circulator into a 100-240 VAC, 47-63 Hz power source.
5. Turn the circulator on. Make sure there are no leaks or air bubbles in the hoses.
6. Turn the camera on.
7. Start WinView.
PIXIS-XB System Manual
If the camera is ON, the problem may be with the USB 2 card, its driver, or the
cable connections.
If the interface card is not installed, close WinView and turn the camera OFF.
Follow the interface card installation instructions provided with the interface
card, and cable the card to the USB 2 port on the rear of the camera. Then do a
"Custom" installation of WinView with the appropriate interface component
selected. Be sure to deselect the interface component that does not apply to your
system.
If the interface card is installed in the computer and is cabled to the USB 2 port
on the rear of the camera, close the application program and turn the camera
OFF. Check the cable connections.
If the interface card was installed after WinView was installed, close WinView
and do a "Custom" installation of WinView/32 with the appropriate interface
component selected. Be sure to deselect the interface component that does not
apply to your system.
and add coolant.
II
reservoir contains coolant. If additional coolant is required, fill with a 50:50 mixture
of water and ethylene glycol.
Note: Small bubbles (about the size of bubbles in soda) are common in the
CoolCUBE
especially at startup and do not prevent proper operation.
II
If there are no problems, continue to Step 6.
If there are leaks or air bubbles, turn the circulator off and correct the problem(s)
by securing the hoses or adding more coolant to the reservoir. Turn the circulator
back on. Recheck and if there are no problems, continue to Step 6.
) and make sure that the coolant
II
Version 2