4
Performance Tests and Calibration
Finishing the Adjustments
108
•
An adjustment failure is indicated by a long beep, the
primary display showing FAiL and a calibration error
number appearing in the secondary display. Check the
input value, range, function, and entered adjustment
value to correct the problem and repeat the adjustment
step.
8 Repeat steps 3 through 7 for each gain adjustment point
shown in the table.
9 Verify the Capacitance Gain adjustments using the
"Optional Capacitance Performance Verification Test"
page 86.
Table 21
Capacitance Gain Adjustment
Input
Input terminals open (remove any test leads or
shorting plugs from the input terminals)
0.4 nF
1 nF
10 nF
100 nF
μ
1
F
μ
10
F
μ
100
F
μ
1000
F
μ
10000
F
1 Remove all shorting plugs and connections from the
instrument.
2 Reset the Calibration Message (see below).
Function
Capacitance Open
34405A User's and Service Guide
on
Cal Item
0.4 nF
1 nF
10 nF
100 nF
μ
1
F
μ
10
F
μ
100
F
μ
1000
F
μ
10000
F