Temperature And Capacitance Specifications [1] - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

Temperature and Capacitance Specifications
Table 26
Temperature and Capacitance Accuracy ± (% of reading + % of range)
Function
Range
Temperature
-80.0° C to 150°C
-110.0° F to 300.0° F
Capacitance
1.000 nF
10.00 nF
100.0 nF
1.000 µF - 100.0 µF
1000 µF
10,000 µF
34405A User's and Service Guide
Probe Type
or Test Current
5 kΩ thermistor probe
5 kΩ thermistor probe
0.75 µA
0.75 µA
8.3 µA
83 µA
0.83 mA
0.83 mA
[1] Specifications are for 30 minutes warm-up, 5½-digit resolution and calibration
temperature 18° C - 28° C.
[2] 20% over range on all ranges except 1000Vdc.
[3] Specifications are 2-wire ohms using Math Null. If without Math Null, add 0.2Ω
additional error.
[4] Specifications are for the voltage measured at the input terminals only.
[5] 20% over range on all ranges except 750 VAC
[6] Specifications are for sinewave inputs > 5% of range. Maximum crest factor : 3
at full scale.
[7] Additional error to be added as frequency > 30kHz and signal input < 10% of
range. 30kHz ~ 100kHz: 0.003% of full scale per kHz.
[8] For 12A terminal, 10A DC or AC rms continuous, > 10A DC or AC rms for 30
seconds ON and 30 seconds OFF.
[9] For 1A and 10A ranges, the frequency is verified for less than 5kHz.
[10] Specifications are for half-hour warm-up, using 0.1 second aperture. The
frequency can be measured up 1Mhz as 0.5V signal to 100mV/1V ranges.
[11] For 20Hz - 10kHz, the sensitivity is AC input current from 10% to 120% of range
except where noted.
[1]
1 Year
23° C ± 5° C
Probe accuracy + 0.2 °C
Probe accuracy + 0.4 °F
2+0.8
1+0.5
1+0.5
1+0.5
1+0.5
2+0.5
Specifications
6
Temperature Coefficient
0° C - 18° C
28° C - 55° C
0.002 °C
0.0036 °F
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001
0.02+0.001
133

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents