Optional AC current performance verification test

Table 2-10 Optional AC current performance verification test

Current
10 mA
10 mA
10 mA
10 mA
1 A
100 mA
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
These tests are not intended to be performed with every
calibration. They are provided as an aid for verifying
additional instrument specifications.
AC current configuration:
(CONFigure:CURRent:AC [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1 Select the AC Current function.
2 Select each range in the order based on the indicated
input current and frequency. Compare the measurement
results with the appropriate test limits shown in the
Table 2-
settling.)
Input frequency
Function
20 Hz
45 Hz
1 kHz
10 kHz
AC Current
1 kHz
1 kHz
[1]
Select NPLC 20 digit resolution.
10. (Be certain to allow for appropriate source
[1]
Range
10 mA
10 mA
10 mA
10 mA
1 A
1 A
Performance Test Calibration
Error from nominal 1
year
±0.16 mA
±0.16 mA
±0.06 mA
±0.22 mA
±6 mA
±1.5 mA
2
25