AC Current Verification Test

Configuration: AC current
1 Select the AC current function.
2 Select each range in the order shown in

Table 4-9 AC current verification test

Connect calibrator to multimeter's A and Lo terminals before applying 10 A.
C A U T I O N
U3402A User's and Service Guide
current and frequency indicated. Compare measurement results to the
appropriate test limits shown in
appropriate source settling.)
Function
Reading
rate
AC current
Slow
Medium
Table 4-
Table 4-
9. (Be certain to allow for
Current
Input
frequency
1.2000 mA
1 kHz
10.0000 mA
1 kHz
12.000 mA
1 kHz
100.000 mA
1 kHz
120.00 mA
1 kHz
1000.00 mA
1 kHz
1.2000 A
1 kHz
10.000 A
1 kHz
4.000 mA
1 kHz
36.000 mA
1 kHz
12.00 mA
1 kHz
100.00 mA
1 kHz
120.0 mA
1 kHz
1000.0 mA
1 kHz
1.200 A
1 kHz
10.000 A
1 kHz
Performance Test
9. Provide the input
Range
Error from nominal
one year
12 mA
±16 µA
12 mA
±60 µA
120 mA
±0.16 mA
120 mA
±0.6 mA
1200 mA
±1.6 mA
1200 mA
±6 mA
12 A
±22 mA
12 A
±110 mA
40 mA
±60 µA
40 mA
±0.22 mA
120 mA
±0.18 mA
120 mA
±0.62 mA
1200 mA
±1.8 mA
1200 mA
±6.2 mA
12 A
±24 mA
12 A
±112 mA
4
73