Starting up
6.4

Menu functions

6.4.1

Refrigeration system (level 1)

10.08.99
YACE0060
or
or
− Switching from setpoint selection (level 0) to level 1 using
the key
− Within one level it is possible to scroll using the keys.
− In principle, after each setting has been made it is entered
automatically after approx. 4 s or
− Settings are entered immediately on operating this key
− The display shows the current bath temperature, COOL and
the set operating mode. To alter the setting, press the key
on the left.
− The display flashes (approx. 4 s). During this time start to
set the operating mode with one of the keys.
0 = Refrigeration system OFF
1 = Refrigeration system ON 100 % capacity
2 = Refrigeration ON, 50% capacity (only RE 207, 212,
220, 307, 312, 320)
A = automatic operation (
− Forward with key to "pump output" or
− with key back to level 0 (setpoint input).
Section 3.6.)