Preparations
5.3

Bath liquids and hose connections

Bath liquids
LAUDA
Designation
Former
designation
water
Kryo 30
G 100
Kryo 50
SK Super
Frigor
Kryo 20
160 MS
Ultra 350
330 SCB
Therm 230
RDS 50
At higher temperatures ! Evaporation losses ! Use bath covers (
Distilled water or fully deionised water must only be used with the addition of 0,1g sodium
carbonate/l water, otherwise
! danger of corrosion!
Water content falls after prolonged operation at higher temperatures
! mixture becomes flammable (flash point 128 °C).
! Check the mixture ratio with a densiometer.
− When selecting bath liquids it should be noted that performance must be expected to worsen
at the lower limit of the operating temperature range due to increasing viscosity. The full
operating range should only be utilised if really necessary.
− The operating ranges of the bath liquids and tubing represent general data which may be
limited by the operating temperature range of the unit.
DIN Safety data sheets are available on request
12.07.00
YACE0060
Working
Chemical
tem-
Designation
perature
range
from °C
at 20°C
to °C
deionised
+5...+90
water
Mono-
-30...+90
ethylene-
glycol/water
Silicone oil
-50...+95
Silicone oil
-20...+180
synthetic
+30...+200
thermal oil
Silicone oil
+60...+230
Silicone oil causes pronounced swelling of Silicone rubber ! never use Silicone oil with
Silicone tubing!
Visco-
Viscosity
sity
(kin) at
(kin)
Temperature
mm²/s
mm²/s
--
--
4
50 at -25°C
6
35 at -50°C
11
28 at -20°C
47
28 at +30°C
44
28 ... +60°C
Fire-
Ref.No.
point
Quantity
5 l
10 l
--
--
LZB 109 LZB 209 LZB 309
> 139
LZB 103 LZB 203 LZB 303
> 230
LZB 116 LZB 216 LZB 316
> 240
LZB 107 LZB 207 LZB 307
> 362
LZB 117 LZB 217 LZB 317
Section 10. Accessories).
20 l
15