Inductance (L) Measuring Procedures - Lutron Electronics LCR-9083 User Manual

Lcr meter
Table of Contents

Advertisement

4. INDUCTANCE (L) MEASURING
PROCEDURES
1) Slide the " Power On/Off switch " (3-2, Fig. 1) to the " 1 "
position.
" 1 " = On
2) Slide the " L/C,R switch " ( 3-4, Fig. 1 ) to the " L/C "
position.
3) Rotate the function switch for the maximum expected
inductance range.
4) Plug the " Test alligator clips " to the " Measuring input "
(3-5, Fig.1), then connect the inductor to the alligator
clips.
5) Read the display. The value indicated corresponds to
the range selected. If the DISPLAY shows "1", it
indicates an Out-of-Range measurement. In order
to improve the resolution, Select the next higher range.
NOTE:
If the inductance value is unmarked start
1) *
with the Lower range ( 2 mH ) and keep
increasing until a suitable reading is obtained.
2)
For the 200 mH range are designed for measuring
*
from 20 mH to 200 mH only. It is normal, if get the no
convineant reading value when measue inductance less
than 20 mH of 200 mH range.
For the 2H range are designed for measuring from 0.2 H
*
to 2H only. It is normal, if get the no convineant
reading value when measue inductance less than 0.2 H
of 2 H range.
For the 20H range are designed for measuring from
*
2 H to 20H only. It is normal, if get the no convineant
reading value when measue inductance less than 2 H
of 20 H range.
" 0 " = Off