SECTION 7—COMPRESSOR
4. Disconnect the spade connectors from the capacitor terminals (not shown).
5. Remove the screw that secures the capacitor clamp to the soundbox brace.
6. Remove the capacitor from the concentrator base assembly.
7. Install new capacitor by reversing STEPS 3-6.
8. Re-install the cabinet. Refer to Removing Cabinet on page 25.

10-Liter Model (Platinum 10)

NOTE: For this procedure, refer to Detail "B" in FIGURE 7.3.
1. Unplug the unit.
2. Remove the cabinet. Refer to Removing Cabinet on page 25.
3. Cut the two tie-wraps that secure the capacitor to the soundbox brace.
4. Rotate capacitor to horizontal position.
5. Remove protective boot to reveal capacitor wires.
6. Disconnect the spade connectors from the capacitor terminals (not shown).
7. Remove the capacitor from the concentrator base assembly.
8. Install new capacitor by reversing STEPS 3-7.
9. Re-install the cabinet. Refer to Removing Cabinet on page 25.
DETAIL "A" - PLATINUM 5 AND
DETAIL "B" - PLATINUM 10 MODELS
PLATINUM XL MODELS
Capacitor
Protective
Boot
Tie-Wraps
Capacitor
Clamp
Screw
Soundbox
Soundbox
Brace
Brace
Protective
Capacitor
Boot
FIGURE 7.3 Replacing Capacitor
36
Platinum™Series
Part No 1110538