Replacing Capacitor; 5-Liter Models (Platinum 5 And Platinum Xl) - Invacare Platinum 5 Service Manual

Platinum series 5 liter oxygen concentrator
Hide thumbs Also See for Platinum 5:
Table of Contents

Advertisement

Platinum 5 and 10 models are
equipped with Thomas Compressors
Mounting
Screws
NOTE: Models with throttling muffler
manufactured after 7/24/03.
FIGURE 7.2 Replacing Compressor Assembly - Platinum 5 and 10 Models with Throttling

Replacing Capacitor

5-Liter Models (Platinum 5 and Platinum XL)

NOTE: For this procedure, refer to Detail "A" in FIGURE 7.3 on page 36.
1. Unplug the unit.
2. Remove the cabinet. Refer to Removing Cabinet on page 25.
3. Remove protective boot to reveal capacitor wires.
Part No 1110538
Intake Hose
Compressor
Restraint Bracket
Muffler
35
SECTION 7—COMPRESSOR
Compressor
Output Hose
(BLUE)
Protective Boot
Platinum 5 XL model is equipped
with Gast Compressor
Grommet
Wire Clamps or
Tie-wraps
Capacitor
Platinum™Series