SECTION 17—ON/OFF SWITCH

Replacing On/Off Switch

To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
NOTE: For this procedure, refer to FIGURE 17.1.
1. Unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 25.
3. Remove control panel. Refer to Removing Control Panel on page 62.
4. Remove and code the four spade connectors from back of existing on/off switch.
5. Compress retaining grips on back of existing on/off switch and push switch out
through front of control panel.
NOTE: Ensure proper orientation of the On/Off (
installation. Refer to FIGURE 17.1, Detail "A" and Detail "B" below.
DO NOT install the On/Off (I/O) switch upside down. Universal Off (O) symbol
should be at bottom and Universal On (I) symbol should be at the top. Possible
damage to the concentrator may result if not properly installed.
6. Reverse STEP 4 to secure new On/Off (
7. Reinstall control panel. Refer to Removing Control Panel on page 62.
8. Reinstall cabinet. Refer to Removing Cabinet on page 25.
Part No 1110538
DANGER
CAUTION
Control Panel
ON/OFF
(
)
Switch
FIGURE 17.1
Replacing On/Off Switch
69
SECTION 17—ON/OFF SWITCH
) switch and wiring connections before
) switch.
DETAIL "A"
ON
OFF
Front View
DETAIL "B"
Spade Connector
from Brown Wire
attached here
Brown Jumper
Wire (From
Circuit Breaker)
Back View
(
)
On/Off
Switch
Spade Connectors
from Black Wires
attached here
Platinum™Series