SECTION 4—TROUBLESHOOTING
SYMPTOM
POWER LOSS:
(Continued)
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch, alarm off. Battery
Free circuit drained
INTERNAL POWER
LOSS SENSO
:
2
Internal Status Indicators:
RED: Off
GREEN: Off
Alarm may or may not be
on.
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Fan operates,
Compressor not
operating.
Platinum™Series
PROBABLE CAUSE
(
)
On/Off
switch.
a. Disconnected wire.
b. Faulty switch.
Circuit breaker tripped.
P.C. board.
a. P.C. board damaged.
b. Loose or damaged connector.
Transformer assembly.
a. Faulty.
b. Connector loose or disconnected.
c. Faulty wiring.
P.C. board faulty.
Check all electrical connections to the
(
ON/OFF
disconnected wires.
If the concentrator does not come on at all
and wiring is intact, color code and remove
wires one at a time. Remove and replace
with new one. Transfer wires from old
switch to new switch one at a time to the
matching contact.
Reset breaker.
NOTE: Breaker may trip to safeguard
concentrator during a power surge.
If breaker trips IMMEDIATELY, there is a
probable short in the unit. Check for
pinched or charred wires. If the breaker
does not trip, run unit for approximately
two hours. The circuit breaker should be
less than 10 ohms. If breaker trips again,
there is an internal problem. Contact
INVACARE TECHNICAL SERVICE at 1-
800-832-4707.
a. Replace P.C. Board. Refer to
Replacing P.C. Board on page 66.
b. Repair or replace connector.
a. Replace. Refer to Replacing
Transformer Assembly on page 68.
b. Reattach connector.
c. Replace transformer assembly. Refer
to Replacing Transformer Assembly
on page 68.
Replace P.C. board. Unit requires retiming
after P.C. board replacement. Refer to
Replacing P.C. Board on page 66.
18
SOLUTION
)
switch for any
Part No 1110538

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Platinum 10Platinum xl

Table of Contents