SECTION 25—SIEVE BED PRESSURE CHECK
SECTION 25— SIEVE BED PRESSURE
CHECK
Installing the Sieve Bed Pressure Tap-In Kit Installation
and Checking Bed Pressure
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.

Disassembly

NOTE: For this procedure, refer to FIGURE 25.1.
1. Turn power off ( ) and unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 25.
3. Turn unit so the back faces you.
4. Loosen and lower the adjustable clamp on the product tank.
5. Pull product tank up and back to access the check valve PVC tubing.
6. Locate the check valve and PVC tubing to your right.
7. Cut the PVC tubing below the check valve in half.
Check Valve
Cut PVC Tubing below
Check Valve to install
Tap-in Kit
Adjustable Clamp
Product Tank
FIGURE 25.1 Disassembly
96
Platinum™Series
Part No 1110538