CONNECTING TO OTHER INSTRUMENTS

Overview

RS-232 Communication Protocol

Connecting the N-7000 Interface
Connecting the N-7500 Network
Connecting the N-9000 Recorder/Interface
Connecting Other Strip-Chart Recorders
Connecting the ThinkJet Printer
Connecting the P-200 Printer
OVERVIEW
WARNING: When connecting this monitor to any
instrument, verify proper operation before clinical use.
Refer to the other deviceÕ s manual for full instructions.
Accessory equipment connected to the monitorÕ s analog
or digital outputs must be certified according to the
respective standards, that is, IEC 950 for data-
processing equipment or IEC 601.1 for electromedical
equipment and must comply with the requirements for
medical systems, IEC 601.1.1. If in doubt, contact
NellcorÕ s Technical Services Department or your local
Nellcor representative.
This section contains information to allow the user to provide
serial communications between the N-200 and external digital
devices.
RS-232 COMMUNICATION PROTOCOL

Serial Data Connector Pin Assignments

Table 4 lists pin assignments on the rear-panel SERIAL
COMM connector and the corresponding pin assignments for a
25-pin RS-232 connector.