Stryker Sidne Operating And Maintenance Manual page 106

Operating-room control system
Table of Contents

Advertisement

Caution Programming the digital board may take several
minutes. To avoid equipment damage, please do not
remove the dongle card before Sidne™ says, "Please
remove card."
When Sidne™ says this...
"Software card detected. Please stand
by."
"Programming 10% complete."
"Programming failed."
"Programming complete."
"Please remove card."
* Sidne™ will announce the completion percentage
throughout the installation process. Do not remove the
dongle until Sidne says, "please remove card."
** Sidne™ will announce this message only if the
programming has failed.
6. Wait as Sidne™ restarts. When the console says, "Sidne™
ready," the driver is installed and ready for use.
Caution Sidne™ should say "ESU Connected" and the EDAM
LED should be flashing green within 20 seconds of
powering On the ESU. A different LED output may
indicate device problems. If the LED does not flash
green after 20 seconds, refer to the "Troubleshooting
Guide" of this addendum for instructions.
7. Power On the ESU.
Please retain this addendum for future reference.
*
**
Installing the ESU Driver Upgrade
do this...
Wait for the next instruction.
Wait for the next instruction.
Remove and reinsert the dongle.
Wait for the next instruction.
Remove the driver dongle from the
Sidne™ console.
101 EN