Settings in the menu
3.7
tESt SC
External
KboArd
PUSH 1 ... 6
dISPLAY
SNr
LiSt
rESEt.AL
SUrE?
28
Diagnosis and printing out of the menu settings (DIAGNOS)
Testing scale with external calibration weight
1. The scale checks the zero point. -0- appears in the display. The test weight
flashes in the display.
2. If necessary, change the weight value displayed with
3. Put the calibration weight on the scale and confirm with
4. The scale checks the calibration weight put on them.
5. After the test is completed, the deviation from the last calibration briefly appears
in the display, ideally *d=0.0g, after which the scale changes to the next menu
item KboArd.
Keyboard test
•
Press the keys
scale changes to the next key.
Note
You cannot abort the keyboard test!
If you have selected the menu item KboArd, you must press all keys.
Display test: The scale displays all functioning segments
Display of the serial number
Printout of a list of all menu settings
Resetting all menu settings to the factory settings
Confirmation inquiry
•
Reset all menu settings to the factory settings with
•
Do not reset the menu settings with